Flow Cytometry (Autofluorescence) severe

Poor Resolution of Dim Markers Masked by Autofluorescence

Symptom
Low-expression markers become indistinguishable from background. Positive and negative populations show poor separation, with dim fluorophores completely masked by cellular autofluorescence.
Common Causes
  1. 1 Dim fluorophores (low-brightness dyes) easily masked by autofluorescence signal
  2. 2 FITC highly susceptible due to overlap with NADH and flavin emissions
  3. 3 UV (355 nm) and blue (488 nm) lasers induce strongest autofluorescence affecting FITC and BUV dyes
  4. 4 Low-expression antigens generating weak signal compared to background
Solutions
  1. 1 Assign bright dyes (PE, BV421) to low-expression markers instead of FITC
  2. 2 Avoid FITC for critical markers; use dyes with emissions outside autofluorescence range
  3. 3 Use red/far-red lasers (633 nm and beyond) which are least affected by autofluorescence
  4. 4 Include FMO (Fluorescence Minus One) controls to distinguish true signal from background
  5. 5 Consider spectral unmixing to separate dim marker signal from autofluorescence
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 72
BD FACSCelesta Multicolor Flow Cytometry Training
"Multicolor hands-on training covers detector/gain setup where autofluorescence masking of dim markers is directly addressed."
Bilibili (China-Accessible Mirrors) ★ 68
Flow Cytometry Complete Workflow: Sample to Analysis
"End-to-end workflow including troubleshooting helps contextualize autofluorescence and dim-signal resolution issues."
BD Biosciences ★ 65
Cell Preparation for Flow Cytometry
"Cell preparation best practices aimed at ensuring proper resolution, relevant to background/autofluorescence problems."
Source: abcam.com ↗
← Back to all cases