Home › Failure Case Library › High Side Scatter Background from Small Particles
Flow Cytometry (Troubleshooting) moderate

High Side Scatter Background from Small Particles

Symptom
Flow cytometry SSC channel shows elevated background noise from small particles and debris. Event plots display excessive scatter in low SSC/FSC regions, indicating presence of cell fragments or contaminants.
Common Causes
  1. 1 Cell lysis due to non-fresh samples or improper preparation protocols, creating membrane fragments
  2. 2 High centrifuge rotor speed or excessive vortexing causing mechanical cell damage and fragmentation
  3. 3 Bacterial contamination in sample producing autofluorescence at low level and high event rates
Solutions
  1. 1 Ensure cells in sample have not lysed; use fresh samples and proper preparation techniques
  2. 2 Do not centrifuge cells at high rotor speed (keep below 300-400 x g); avoid violent vortexing; handle cells gently throughout protocol
  3. 3 Ensure sample is not contaminated; work in sterile conditions; use aseptic technique during cell preparation; filter samples if necessary
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 80
Flow Cytometry Experimental Operation in 7 Minutes
"Covers sample preparation and instrument operation, directly addressing the prep-related lysis cause behind high SSC debris background."
Bilibili (China-Accessible Mirrors) ★ 78
Flow Cytometry Complete Workflow: Sample to Analysis
"Includes troubleshooting within the sample-to-analysis workflow, relevant to diagnosing and preventing sample prep artifacts like fragmentation."
Bilibili (China-Accessible Mirrors) ★ 70
Ten-Minute Guide to Flow Cytometer Instrument Operation
"Demonstrates scatter plots and parameter settings, useful for recognizing and interpreting low SSC/FSC debris on event plots."
Source: abcam.com ↗
← Back to all cases