Home Failure Case Library Photobleaching During Fixation and Storage
Flow Cytometry (Fixation & Permeabilization) moderate

Photobleaching During Fixation and Storage

Symptom
Progressive loss of fluorescence signal over time, particularly noticeable in samples fixed and stored for later analysis. Light-sensitive fluorophores show dramatically reduced intensity.
Common Causes
  1. 1 Samples exposed to ambient light during fixation and permeabilization steps
  2. 2 Inadequate light protection during overnight storage of fixed cells
  3. 3 Photobleaching of fluorophores during extended processing times
  4. 4 Use of clear tubes instead of light-protected containers
Solutions
  1. 1 Keep samples protected from light during fixation and all subsequent steps
  2. 2 Use aluminum foil or amber tubes to shield samples from light exposure
  3. 3 Minimize time between staining and acquisition, even for fixed samples
  4. 4 Store fixed cells at 4°C in darkness if overnight storage is required
Related Video (3)
BioLegend ★ 85
Surface and Intracellular Cytokine Staining for Flow Cytometry
"Directly demonstrates fixation and permeabilization for flow cytometry, where light exposure during these steps causes photobleaching."
Bilibili (China-Accessible Mirrors) ★ 70
Flow Cytometry Experimental Operation in 7 Minutes
"Shows full flow cytometry protocol including sample prep and handling, useful for identifying where light exposure occurs."
Cell Signaling Technology ★ 50
Formaldehyde vs. alcohol fixation for immunofluorescence (IF) | CST Tech Tips
"Explains fixation methods for fluorescence, but focused on IF not flow cytometry, moderately helpful for understanding fixative effects on fluorophores."