Home Failure Case Library Fluorescence Spillover into Secondary Detectors
Flow Cytometry (Controls) severe

Fluorescence Spillover into Secondary Detectors

Symptom
One fluorochrome's emission spectra spills over into another detector channel, creating false positive signals. Data appears contaminated with signals that do not represent true marker expression.
Common Causes
  1. 1 Insufficient compensation applied to correct for spectral overlap between fluorochromes
  2. 2 Single-stain controls not properly measured for each fluorochrome in the panel
  3. 3 Compensation calculated incorrectly or not applied during acquisition or analysis
  4. 4 Multicolour panel design with fluorochromes having significant spectral overlap
Solutions
  1. 1 Prepare single-stain controls using beads or cell samples for each fluorescent marker in the panel
  2. 2 Use software to calculate and apply correct compensation either in real time during acquisition or afterward during analysis
  3. 3 Run compensation controls before each multicolour experiment to determine degree of spillover
  4. 4 Optimize panel design to minimize spectral overlap between fluorochromes when possible
Related Video (3)
BD Biosciences ★ 95
Flow Cytometry Compensation Tips and Tricks
"Directly addresses compensation to correct spectral overlap, the exact cause of the failure."
Bilibili (China-Accessible Mirrors) ★ 60
BD FACSCelesta Software Operation Training
"Shows software setup including detector and compensation settings on a common BD instrument."
Bilibili (China-Accessible Mirrors) ★ 55
Flow Cytometry Experimental Operation in 7 Minutes
"General protocol covers instrument parameter adjustment, providing context for controlling fluorescence detection."
Source: bdbiosciences.com ↗
← Back to all cases