Incorrect Positive/Negative Cell Population Ratios

Symptom
The measured ratio of positive to negative cells for a given marker appears inaccurate or inconsistent. Background signals are not correctly measured, leading to improper gating and incorrect population quantification.
Common Causes
  1. 1 Improper controls used, failing to account for different sources of background signal
  2. 2 Incorrect instrument gain (photomultiplier tube voltages) settings not optimized for the fluorochromes used
  3. 3 Laser alignment, laser time delay, or sensitivity issues not checked periodically
  4. 4 Inappropriate gate boundaries set between positive and negative expression events
Solutions
  1. 1 Use appropriate experiment-associated controls (single-stain, FMO, isotype, or biological) to identify main source of background
  2. 2 Set instrument gain properly for multicolour experiments by measuring spillover and determining required compensation
  3. 3 Perform periodic quality control checks of instrument optics, electronics and fluidics using BD CS&T beads with BD FACSDiva Software
  4. 4 Use biological controls (e.g., unstimulated samples) to set relevant boundaries between positive and negative expression when appropriate
Related Video (3)
BD Biosciences ★ 85
Flow Cytometry Compensation Tips and Tricks
"Directly addresses compensation and control strategies in flow cytometry, essential for proper background signal measurement and gating accuracy."
BD Biosciences ★ 78
Cell Preparation for Flow Cytometry
"Covers cell preparation best practices that impact background signal quality and proper resolution needed for accurate positive/negative population gating."
Bilibili (China-Accessible Mirrors) ★ 72
Flow Cytometry Complete Workflow: Sample to Analysis
"Comprehensive workflow including troubleshooting section likely covers control setup and gating strategies relevant to correcting population ratio errors."
Source: bdbiosciences.com ↗
← Back to all cases