Home Failure Case Library Endogenous Biotin Interference
Immunohistochemistry (CST Guide) moderate

Endogenous Biotin Interference

Symptom
High background staining specifically in kidney, liver, or other biotin-rich tissues when using biotin-based detection. Background not resolved by standard blocking procedures.
Common Causes
  1. 1 Biotin-based detection system used on tissues with high endogenous biotin levels (kidney, liver)
  2. 2 Endogenous biotin binds avidin/streptavidin detection reagents creating false-positive signal
  3. 3 No biotin blocking step performed prior to primary antibody incubation
Solutions
  1. 1 Switch to polymer-based detection system (SignalStain Boost IHC Detection Reagents #8114 or #8125) which eliminates biotin interference
  2. 2 If biotin-based detection must be used, perform biotin block after normal blocking procedure and prior to primary antibody incubation
  3. 3 Avoid biotin-based systems entirely for kidney and liver tissue IHC experiments
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 90
Immunohistochemistry on Paraffin-Embedded Sections (CST Demo)
"Demonstrates CST's paraffin IHC workflow, helpful for seeing alternative detection systems to avoid biotin interference."
Cell Signaling Technology ★ 85
Immunohistochemistry Protocol for Paraffin embedded Tissue Sections
"Shows proper IHC protocol on paraffin sections from CST, including detection steps, to contrast with biotin-based methods."
Cell Signaling Technology ★ 70
Better IHC Step 1: Antigen Retrieval
"Focuses on improving IHC staining quality, relevant to troubleshooting background issues in biotin-rich tissues."
Source: cellsignal.com ↗
← Back to all cases