Immunohistochemistry (CST Guide) moderate

Inadequate Washing Steps Causing Poor Contrast

Symptom
High background persists with poor contrast between positive signal and background despite other optimization steps, making interpretation difficult.
Common Causes
  1. 1 Washing slides fewer than 3 times after antibody incubations
  2. 2 Washing duration less than 5 minutes per wash step
  3. 3 Using incorrect wash buffer (not 1X TBST #9997)
  4. 4 Insufficient agitation or mixing during wash steps
Solutions
  1. 1 Wash slides exactly 3 times for 5 minutes each with 1X TBST (#9997) after primary antibody incubation
  2. 2 Wash slides exactly 3 times for 5 minutes each with 1X TBST (#9997) after secondary antibody incubation
  3. 3 Use only recommended 1X TBST (#9997) as wash buffer; prepare fresh if needed
  4. 4 Ensure adequate agitation during washing to remove unbound antibodies effectively
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 88
Immunohistochemistry on Paraffin-Embedded Sections (CST Demo)
"CST-based step-by-step paraffin IHC demo covers washing and incubation steps where inadequate washing causes high background."
Bilibili (China-Accessible Mirrors) ★ 75
Complete IHC Immunohistochemistry Protocol for Beginners
"Complete beginner-friendly IHC protocol demonstrates the full workflow including washing steps between antibody incubations."
Cell Signaling Technology ★ 72
Immunohistochemistry Protocol for Paraffin embedded Tissue Sections
"CST IHC protocol video directly matches the guide source and shows correct washing/incubation procedure to avoid poor contrast."
Source: cellsignal.com ↗
← Back to all cases