Home › Failure Case Library › Sequence Errors Within PCR Product Body
PCR (Invitrogen Guide) severe

Sequence Errors Within PCR Product Body

Symptom
Sequencing reveals point mutations, insertions, or deletions within the amplified fragment that are not present in original template.
Common Causes
  1. 1 Low-fidelity DNA polymerase lacking proofreading activity
  2. 2 Excess Mg2+ concentration favoring nucleotide misincorporation
  3. 3 Unbalanced dNTP concentrations (non-equimolar dATP, dCTP, dGTP, dTTP) increasing error rate
  4. 4 High number of PCR cycles increasing mismatched nucleotide incorporation
  5. 5 UV damage from short-wavelength (254–312 nm) light box exposure
Solutions
  1. 1 Use DNA polymerases with exceptionally high fidelity for cloning, sequencing, and mutagenesis applications
  2. 2 Review and reduce Mg2+ concentration to prevent misincorporation
  3. 3 Ensure equimolar concentrations (typically 200 µM each) of dATP, dCTP, dGTP, and dTTP
  4. 4 Reduce cycle number without lowering yield; increase template input to avoid excessive cycles
  5. 5 Use long-wavelength UV (360 nm) light box with minimal illumination time; alternatively use longer-wavelength excitation dyes
  6. 6 Sequence both DNA strands with duplicate samples to verify results
Related Video (2)
Addgene ★ 72
Polymerase Chain Reaction (PCR) Protocol
"Addgene PCR protocol walkthrough shows the core PCR workflow where polymerase choice is made, the direct context for this fidelity failure."
Bilibili (China-Accessible Mirrors) ★ 68
Step-by-Step PCR Experiment Protocol Guide
"Comprehensive hands-on PCR protocol demonstration covers setup and result analysis, giving context for where low-fidelity polymerase errors arise."
Source: thermofisher.com ↗
← Back to all cases