Home Failure Case Library Residual Adaptor or Primers After PCR
NGS Library Preparation (NEBNext Ultra II) moderate

Residual Adaptor or Primers After PCR

Symptom
Adaptor or primer peaks visible on Bioanalyzer or similar instrument after PCR amplification, indicating inefficient cleanup or excess reagent usage.
Common Causes
  1. 1 Excess adaptor used during ligation step relative to sample input
  2. 2 Excess PCR primer added or inefficient post-PCR cleanup
  3. 3 SPRI bead cleanup ratio insufficient to remove small adaptor/primer fragments
Solutions
  1. 1 Perform another 0.9X SPRI bead cleanup to remove residual adaptors and primers
  2. 2 Optimize adaptor concentration during ligation based on input amount
  3. 3 Verify correct PCR primer volume is being added according to manual specifications
Related Video (2)
New England Biolabs ★ 95
NEBNext Ultra II DNA Library Prep Protocol
"Direct protocol walkthrough for NEBNext Ultra II DNA Library Prep Kit, the exact technique experiencing the failure"
New England Biolabs ★ 78
12 Quick Tips for NGS Library Preparation
"Covers 12 quick tips for NGS library prep including adaptor handling and cleanup best practices relevant to this failure mode"
Source: neb.com ↗
← Back to all cases