Symptom
Low concentration data points do not fit linear standard curve profile; NTC shows amplification with lower Tm and broader melt peak than positive samples; primer dimers visible on gel, inversely proportional to template concentration
Common Causes
  1. 1 Primers are self-dimerizing in absence or at low levels of template
  2. 2 Excessive primer concentration in reaction
  3. 3 Mg2+ concentration too high
  4. 4 Annealing temperature too low allowing non-specific primer interactions
Solutions
  1. 1 Try lower primer concentrations (test 100 nM, 500 nM, 900 nM)
  2. 2 Optimize or re-design the assay to eliminate primer-dimer formation
  3. 3 Titrate Mg2+ to determine optimum concentration
  4. 4 Perform annealing temperature gradient (55°C to 70°C) to select optimal Ta
  5. 5 With RT-PCR, use less RT enzyme, adopt 2-step protocol, or use higher incubation temperature for RT step
Related Video (3)
Thermo Fisher Scientific ★ 90
How to Optimize qPCR using SYBR Green Assays - Ask TaqMan #38
"Directly addresses qPCR optimization and SYBR Green assays, including troubleshooting primer dimers and non-specific amplification."
YouTube (Curated Tutorials) ★ 85
The Features Of A Good qPCR Primer Pair
"Focuses on features of good qPCR primer pairs, which relates to avoiding self-dimerization and primer dimer formation."
YouTube (Curated Tutorials) ★ 75
Primer Design: Important Considerations and Tips for Good Primer Design
"Discusses important primer design considerations and tips to prevent primer dimer issues at low template concentrations."
Source: sigmaaldrich.com ↗
← Back to all cases