Home Failure Case Library Additional Bands or Wrong Molecular Weight
Western Blot Immunodetection moderate

Additional Bands or Wrong Molecular Weight

Symptom
Blot shows unexpected extra bands at different molecular weights than predicted, or expected band appears at wrong size. Multiple bands may appear where single band is anticipated.
Common Causes
  1. 1 Nonspecific primary antibody cross-reacting with off-target proteins
  2. 2 Protein degradation due to insufficient protease inhibitor concentration or sample preparation at room temperature
  3. 3 Post-translational modifications (glycosylation, phosphorylation, ubiquitination) altering protein molecular weight
  4. 4 Protein cleavage generating fragments or isoforms of different sizes
Solutions
  1. 1 Verify antibody specificity using knockout/knockdown cell lysates or tissue lacking target protein; use alternative validated antibody
  2. 2 Prepare cell lysates on ice; add fresh protease inhibitor cocktail and phosphatase inhibitors immediately before lysis
  3. 3 Consult published literature and protein databases for known post-translational modifications affecting molecular weight
  4. 4 Run positive control samples with known protein size; use deglycosylation enzymes (PNGase F) if glycosylation suspected
Related Video (3)
Cell Signaling Technology ★ 92
Western Blot Troubleshooting Guide
"Specifically addresses troubleshooting Western blot problems including extra bands and nonspecific antibody issues."
Bilibili (China-Accessible Mirrors) ★ 68
Western blot hands-on: Protein extraction through detection
"Hands-on protocol highlighting common pitfalls which include antibody-related artifacts leading to additional bands."
Bio-Rad Laboratories ★ 55
Western Blotting
"Shows standard WB procedure including antibody incubation steps where cross-reactivity often originates."
Source: sigmaaldrich.com ↗
← Back to all cases