🚨 Failure Case Library (6) + Submit your own case

critical
Complete Absence of Standard Curve Signal with Normal Zero
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
💡 4 · ✓ 4
severe
Standard Curve Plateau at High Concentration Range
The standard curve reaches a plateau (flat response) at high standard concentrations with correspondingly low OD values, failing to show expected dose-response relationship in competitive ELISA format.
💡 4 · ✓ 4
severe
High CV from Inconsistent Pipetting Technique
High intra-assay coefficient of variation (>15%) between technical replicates. Standard curve shows poor reproducibility between duplicate points.
💡 5 · ✓ 5
severe
High CV from Plate Edge Effects
Outer (edge) wells show systematically different absorbance values compared to inner wells, with CV >15%. Edge wells may show higher or lower signal depending on evaporation or temperature variation. Visual inspection from the side reveals lower buffer levels in edge wells.
💡 5 · ✓ 5
severe
Inadequate Blocking of Non-Specific Binding Sites
Uniformly elevated background across plate with high coefficient of variation. Background may be particularly high in wells with lower antigen concentrations.
💡 4 · ✓ 4
moderate
Over-Washing Removes Bound Detection Reagents
Signal progressively decreases with increased wash steps or aggressive washing. Reducing wash stringency restores signal. Edge wells show lower signal than center wells.
💡 4 · ✓ 5