High Signal in Negative Cell Populations

Symptom
Off-target cell populations such as monocytes show unexpectedly high fluorescence signal. Non-specific staining obscures true positive populations.
Common Causes
  1. 1 Fc surface receptors on monocytes or other cells binding Fc portion of primary or secondary antibodies
  2. 2 Excessive primary or secondary antibody causing non-specific binding
  3. 3 Incomplete washing between antibody incubation steps leaving unbound antibody
Solutions
  1. 1 Block cells with Bovine Serum Albumin, Fc receptor blocking reagents, or normal serum from same host as primary/secondary antibody prior to staining
  2. 2 Include secondary-antibody-only control to determine if background originates from primary or secondary antibody
  3. 3 Perform additional wash steps between antibody incubations to remove unbound antibody