Home Failure Case Library Poor Standard Curve Discrimination
ELISA Development moderate

Poor Standard Curve Discrimination

Symptom
Standard curve achieved but shows low or flat curve with poor discrimination between concentration points
Common Causes
  1. 1 Insufficient streptavidin-HRP
  2. 2 Capture antibody did not bind well to plate
  3. 3 Insufficient detection antibody
  4. 4 Plate not developed long enough
  5. 5 Incorrect procedure followed
  6. 6 Improper calculation of standard curve dilutions
Solutions
  1. 1 Check streptavidin-HRP dilution and titrate if necessary
  2. 2 Use proper ELISA plates (not tissue culture plates)
  3. 3 Dilute capture antibody in PBS without additional protein
  4. 4 Check detection antibody dilution and titrate if necessary
  5. 5 Increase substrate solution incubation time
  6. 6 Use recommended brand of substrate solution
  7. 7 Return to general ELISA protocol and eliminate modifications
  8. 8 Check calculations and make new standard curve
Related Video (3)
Thermo Fisher Scientific ★ 85
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Provides the correct ELISA protocol including streptavidin-HRP steps, helping troubleshoot low signal due to insufficient reagent."
Thermo Fisher Scientific ★ 75
How to coat your own plate and run an Invitrogen ELISA kit
"Covers plate coating and running an Invitrogen ELISA where streptavidin-HRP is used, useful for reviewing the affected steps."
Bilibili (China-Accessible Mirrors) ★ 70
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"Hands-on sandwich ELISA protocol demonstrating reagent addition steps, directly relevant to understanding assay setup and potential causes of poor standard curve."
Source: rndsystems.com ↗
← Back to all cases