Home Immunology How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
Steps
  1. 1 Unpack kit and prepare plate 00:02
  2. 2 Prepare wash buffer and standards 00:48
  3. 3 Perform serial dilution of standard 01:24
  4. 4 Add standards, samples, and detection antibody 01:47
  5. 5 Wash plate and add streptavidin-HRP 02:33
  6. 6 Wash plate and add chromogenic substrate 03:36
  7. 7 Stop reaction and read absorbance 03:54
Immunology Thermo Fisher Scientific

How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial

Protocol
Difficulty
intermediate

Steps

1
Unpack kit and prepare plate

Remove the pre-coated ELISA plate from the foil pouch and separate any unused 8-well strips. Unused strips can be resealed and stored at 2-8°C for future use.

▶ 00:02
2
Prepare wash buffer and standards

Allow wash buffer concentrate to reach room temperature and mix to dissolve any precipitated salts, then dilute with deionized water. Reconstitute protein standard vial by gentle swirling or brief vortexing and allow to sit for at least 10 minutes.

▶ 00:48
3
Perform serial dilution of standard

Create a 1:2 serial dilution by adding assay diluent to each tube, then transferring reconstituted protein standard through the tubes. Mix thoroughly by pipetting up and down, changing tips between tubes.

▶ 01:24
4
Add standards, samples, and detection antibody

Prepare SA buffer and dilute biotin conjugate concentrate. Add diluted biotin-conjugated detector antibody to the plate with standards, controls, and samples, then seal and incubate the plate.

▶ 01:47
5
Wash plate and add streptavidin-HRP

Wash the plate manually or with automated washer by filling with buffer, decanting, and repeating according to protocol. Dilute streptavidin-HRP concentrate in assay buffer, add to plate, and incubate.

▶ 02:33
6
Wash plate and add chromogenic substrate

Wash the plate again to remove excess enzyme. Add chromogenic substrate and incubate at room temperature in the dark for 30 minutes to develop color.

▶ 03:36
7
Stop reaction and read absorbance

Add stop solution to terminate the enzymatic reaction, which turns the wells yellow. Measure absorbance at 450 nanometers using a plate reader.

▶ 03:54

🚨 Failure Case Library (24) + Submit your own case

critical
Absent Standard Curve Signal with Normal Zero Standard
No detectable signal is observed for standard curve points containing analyte, while the zero standard (maximum binding control) shows normal OD values, indicating complete loss of competitive inhibition.
💡 4 · ✓ 4
critical
No Signal or Weak Signal in ELISA
After developing the ELISA plate, wells show no color change or very faint signal that is barely above background, even for positive controls or high-concentration standards.
💡 6 · ✓ 6
critical
No Signal for Standard Curve with Normal Zero Standard
All standard concentrations show no signal (high OD in competitive format similar to zero standard), while the zero standard (no competitor) shows normal maximum binding signal.
💡 4 · ✓ 4
critical
Complete Absence of Standard Curve Signal with Normal Zero
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
💡 4 · ✓ 4
critical
Complete Signal Absence
No detectable signal across the entire plate when signal is expected
💡 7 · ✓ 9
severe
Insufficient Signal from Zero Standard (B0)
The zero standard (no competing antigen, maximum antibody binding) produces OD values that are too low, resulting in compressed standard curve with poor sensitivity.
💡 4 · ✓ 4
severe
Standard Curve Plateau at High Concentration Range
The standard curve reaches a plateau (flat response) at high standard concentrations with correspondingly low OD values, failing to show expected dose-response relationship in competitive ELISA format.
💡 4 · ✓ 4
severe
Standard Curve Plateau at High Concentration End
The standard curve reaches a plateau at high analyte concentrations (low OD values in competitive format), showing no further decrease in signal despite increasing standard concentration.
💡 4 · ✓ 4
severe
OD Value of Zero Standard Too Low
The zero standard (maximum binding, no competitor) shows insufficient optical density signal, providing inadequate dynamic range for the competitive assay and poor B/B0 calculations.
💡 4 · ✓ 4
severe
High CV from Inconsistent Pipetting Technique
High intra-assay coefficient of variation (>15%) between technical replicates. Standard curve shows poor reproducibility between duplicate points.
💡 5 · ✓ 5
severe
Standard Curve R² Below 0.98 Threshold
The regression coefficient (R²) of the standard curve falls below 0.98, indicating poor curve fit and unreliable quantification of sample concentrations.
💡 5 · ✓ 6
severe
Excessive Signal - Entire Plate Turned Blue
Whole plate turned uniformly blue indicating non-specific signal saturation
💡 5 · ✓ 5
severe
B/B0 Ratio at Curve Endpoint Outside 5-95% Range
The B/B0 ratio (bound/maximum bound signal) for the highest or lowest standard concentration exceeds 95% or falls below 5%, indicating the standard curve dynamic range does not adequately cover the analyte concentration range.
💡 4 · ✓ 4
severe
Poor Assay-to-Assay Reproducibility
Inconsistent results between different assay runs with same samples
💡 7 · ✓ 9
moderate
High CV from Bubbles in Wells
Inconsistent absorbance readings between replicate wells with coefficient of variation >15%. Visual inspection may reveal bubbles present in wells, particularly when using detergent-containing solutions.
💡 4 · ✓ 4
moderate
High Background Signal
Elevated background noise across the ELISA plate making it difficult to distinguish specific signals from non-specific binding
💡 3 · ✓ 3
moderate
Unexpected or Inconsistent Assay Results
Sample quantification values are outside expected biological range, controls fail to meet acceptance criteria, or results are not reproducible between runs despite similar experimental conditions.
💡 5 · ✓ 5
moderate
Poor Duplicate Reproducibility
Duplicate wells showing inconsistent results with high variation between replicates
💡 6 · ✓ 8
moderate
Very Low Readings Across Plate
Consistently low optical density readings across entire plate including standards
💡 5 · ✓ 6
moderate
Signal Drift Across Plate
Gradual change in signal intensity across the plate suggesting time-dependent variation during assay setup
💡 2 · ✓ 2
moderate
Weak Signal from Inadequate Antibody-Antigen Interaction
Signal is present but consistently lower than expected. Positive controls show weak but detectable signal while samples are near background.
💡 4 · ✓ 4
moderate
B/B0 Ratio at Standard Curve Endpoint Out of Range
The B/B0 ratio (bound/maximum bound) at the lowest or highest standard concentration is either >95% or <5%, indicating the standard curve dynamic range is not properly positioned.
💡 3 · ✓ 3
moderate
Poor Standard Curve Discrimination
Standard curve achieved but shows low or flat curve with poor discrimination between concentration points
💡 6 · ✓ 8
minor
Standard Curve OD Values Differ From Datasheet
The optical density (OD) measurement values for the standard curve vary considerably from the examples shown on the kit datasheet or protocol booklet, causing user concern about assay validity.
💡 4 · ✓ 4
💬 Comments coming soon