ELISA Development moderate

Very Low Readings Across Plate

Symptom
Consistently low optical density readings across entire plate including standards
Common Causes
  1. 1 Incorrect wavelengths used in plate reader
  2. 2 Insufficient development time
  3. 3 Coated plates are old and degraded
  4. 4 Capture antibody did not bind to plate
  5. 5 Buffer containing FCS used to reconstitute antibodies
Solutions
  1. 1 Check filters and reader wavelength settings
  2. 2 Increase development time
  3. 3 Coat new plates
  4. 4 Use proper ELISA plates (not tissue culture plates)
  5. 5 Dilute capture antibody in PBS without additional protein
  6. 6 Requalify reagents if FCS was used