Home › Failure Case Library › Complete Signal Absence
ELISA Development critical

Complete Signal Absence

Symptom
No detectable signal across the entire plate when signal is expected
Common Causes
  1. 1 Reagents added in incorrect order or improperly prepared
  2. 2 HRP contamination with azide
  3. 3 Insufficient antibody concentration used
  4. 4 Degraded or mishandled standard
  5. 5 Buffer containing FCS used to reconstitute antibodies
  6. 6 Capture antibody failed to bind to plate
  7. 7 Contaminated buffers
Solutions
  1. 1 Repeat assay and review protocol carefully
  2. 2 Check all calculations and prepare fresh buffers and standards
  3. 3 Use fresh reagents free from contamination
  4. 4 Increase antibody concentration
  5. 5 Use new vial of standard handled according to directions
  6. 6 Requalify reagents if FCS was used
  7. 7 Use proper ELISA plates (not tissue culture plates)
  8. 8 Dilute capture antibody in PBS without additional protein
  9. 9 Make fresh buffers
Related Video (3)
Thermo Fisher Scientific ★ 88
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Full step-by-step Invitrogen ELISA protocol demonstrates correct reagent order and preparation, directly addressing the top cause of no signal."
Bilibili (China-Accessible Mirrors) ★ 82
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"Hands-on sandwich ELISA protocol showing coating, antibody steps, and substrate development where ordering/prep errors cause signal absence."
Thermo Fisher Scientific ★ 62
How to coat your own plate and run an Invitrogen ELISA kit
"Shows plate coating and running an Invitrogen ELISA kit, relevant to reagent preparation steps that can go wrong before signal development."
Source: rndsystems.com ↗
← Back to all cases