Symptom
PCR efficiency calculated from standard curve is greater than 120%; ΔCq between 10-fold dilutions is much less than expected 3.3 cycles (e.g., 1.5 cycles); standard curve gradient indicates abnormally high efficiency
Common Causes
  1. 1 Excessive primer-dimer formation particularly in low concentration samples
  2. 2 Pipetting is inaccurate causing errors in serial dilutions
  3. 3 Inhibition present in samples
  4. 4 Low concentration samples contain additional signal from primer dimers
Solutions
  1. 1 Evaluate dissociation/melt curve plot to identify primer dimers (look for lower Tm peak in low concentration samples)
  2. 2 Try lower primer concentrations to reduce primer-dimer
  3. 3 Test pipette calibration and ensure accurate pipetting technique
  4. 4 Check ΔCq between dilutions - if decreasing, indicates inhibitor (confirm with SPUD assay)
  5. 5 Purify sample or use higher dilution in the assay
  6. 6 With RT-PCR: use less RT enzyme, adopt 2-step protocol, or use higher RT incubation temperature
Related Video (3)
Thermo Fisher Scientific ★ 88
How to Optimize qPCR using SYBR Green Assays - Ask TaqMan #38
"Directly addresses qPCR optimization with SYBR Green assays, the exact system where primer-dimer-driven efficiency >120% and abnormal ΔCq arise."
YouTube (Curated Tutorials) ★ 78
The Features Of A Good qPCR Primer Pair
"Explains features of a good qPCR primer pair, helping researchers understand and avoid primers prone to dimer formation."
Bilibili (China-Accessible Mirrors) ★ 72
qPCR Principles, Experimental Workflow and Results Analysis
"Covers qPCR principles and results analysis, giving context for interpreting efficiency and ΔCq standard curve anomalies."
Source: sigmaaldrich.com ↗
← Back to all cases