Home Failure Case Library PCR ReadyMix Works for PCR but Fails in qPCR
PCR (Sigma Guide) severe

PCR ReadyMix Works for PCR but Fails in qPCR

Symptom
ReadyMix produces amplification in standard PCR but completely fails in real-time qPCR applications; equivalent products from other suppliers work well
Common Causes
  1. 1 REDTaq ReadyMix was used for real-time qPCR (red dye interferes with fluorescence detection)
  2. 2 ReadyMix without reference dye was used on instrument requiring reference dye or incorrect concentration was included
  3. 3 ReadyMix designed for use with probes was used without including a probe in the reaction
Solutions
  1. 1 Use appropriate ReadyMixes for the desired PCR application
  2. 2 Ensure red dye products are not used for qPCR as they interfere with fluorescence detection in most qPCR instruments
  3. 3 Verify ReadyMix choice matches instrument requirements and application type
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
qPCR Principles, Experimental Workflow and Results Analysis
"Directly explains qPCR principles and workflow, helping researcher understand why red dye interferes with fluorescence detection."
Thermo Fisher Scientific ★ 80
How to Optimize qPCR using SYBR Green Assays - Ask TaqMan #38
"Discusses qPCR optimization with intercalating dyes, directly addressing dye-related failure in qPCR."
Bilibili (China-Accessible Mirrors) ★ 75
Bio-Rad qPCR Instrument Operation Protocol
"Shows qPCR instrument operation, making clear the fluorescence detection step where the red dye causes failure."
Source: sigmaaldrich.com ↗