High Coefficient of Variation at Low Standard Concentrations

Symptom
Replicate wells at the bottom of the standard curve (low analyte, high OD) show high CVs >15-20%, while mid-curve and high-concentration points demonstrate acceptable reproducibility.
Common Causes
  1. 1 Higher OD values inherently more susceptible to pipetting and reader variability
  2. 2 Minimal competitive displacement at low concentrations amplifies measurement variance
  3. 3 Sample concentrations falling outside optimal middle range of standard curve
  4. 4 Edge effects on microplate causing differential evaporation or temperature gradients
Solutions
  1. 1 Shift standard curve concentration range upward to move samples into mid-curve range with lower CVs
  2. 2 Dilute samples less (e.g., 1:2 instead of 1:5) to increase measured concentration into optimal range
  3. 3 Use calibrated multichannel pipettes and increase number of replicates from duplicates to triplicates
  4. 4 Avoid using outer edge wells of plate or include blank rows/columns as thermal buffer zones
Related Video (2)
Bilibili (China-Accessible Mirrors) ★ 78
Thermo Fisher Microplate Reader Software Operation
"Directly addresses microplate reader software operation and data acquisition, critical for understanding OD measurement variability and CV calculation at high absorbance values."
Bilibili (China-Accessible Mirrors) ★ 72
How to Run an R&D Systems Quantikine ELISA
"Complete Quantikine ELISA workflow with troubleshooting guidance provides practical context for identifying and mitigating high CV issues at low standard concentrations."
Source: abcam.com ↗
← Back to all cases