Home Failure Case Library High Coefficient of Variation at Low Standard Concentrations
ELISA (Competitive) moderate

High Coefficient of Variation at Low Standard Concentrations

Symptom
Replicate wells at the bottom of the standard curve (low analyte, high OD) show high CVs >15-20%, while mid-curve and high-concentration points demonstrate acceptable reproducibility.
Common Causes
  1. 1 Higher OD values inherently more susceptible to pipetting and reader variability
  2. 2 Minimal competitive displacement at low concentrations amplifies measurement variance
  3. 3 Sample concentrations falling outside optimal middle range of standard curve
  4. 4 Edge effects on microplate causing differential evaporation or temperature gradients
Solutions
  1. 1 Shift standard curve concentration range upward to move samples into mid-curve range with lower CVs
  2. 2 Dilute samples less (e.g., 1:2 instead of 1:5) to increase measured concentration into optimal range
  3. 3 Use calibrated multichannel pipettes and increase number of replicates from duplicates to triplicates
  4. 4 Avoid using outer edge wells of plate or include blank rows/columns as thermal buffer zones
Related Video (3)
Thermo Fisher Scientific ★ 85
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Shows proper pipetting and plate-reading technique from a reputable kit manufacturer, directly addressing variability at low concentrations."
Bilibili (China-Accessible Mirrors) ★ 75
Thermo Fisher Microplate Reader Software Operation
"Focuses on microplate reader operation and data acquisition, key to understanding reader-related variability in higher OD wells."
Bilibili (China-Accessible Mirrors) ★ 70
How to Run an R&D Systems Quantikine ELISA
"Includes troubleshooting guidance within a full practical ELISA workflow, helping identify and reduce pipetting-induced CV issues."
Source: abcam.com ↗
← Back to all cases