High Coefficient of Variation at Low Standard Concentrations

Symptom
Replicate wells at the bottom of the standard curve (low analyte, high OD) show high CVs >15-20%, while mid-curve and high-concentration points demonstrate acceptable reproducibility.
Common Causes
  1. 1 Higher OD values inherently more susceptible to pipetting and reader variability
  2. 2 Minimal competitive displacement at low concentrations amplifies measurement variance
  3. 3 Sample concentrations falling outside optimal middle range of standard curve
  4. 4 Edge effects on microplate causing differential evaporation or temperature gradients
Solutions
  1. 1 Shift standard curve concentration range upward to move samples into mid-curve range with lower CVs
  2. 2 Dilute samples less (e.g., 1:2 instead of 1:5) to increase measured concentration into optimal range
  3. 3 Use calibrated multichannel pipettes and increase number of replicates from duplicates to triplicates
  4. 4 Avoid using outer edge wells of plate or include blank rows/columns as thermal buffer zones
Related Video (3)
Thermo Fisher Scientific ★ 78
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Invitrogen step-by-step tutorial emphasizes accurate, reproducible protein detection, directly addressing pipetting/reader variability that drives low-end standard CVs."
Bilibili (China-Accessible Mirrors) ★ 72
Thermo Fisher Microplate Reader Software Operation
"Microplate reader software/operation video helps researchers understand data acquisition and instrument variability, relevant to high-OD reproduction problems."
Bilibili (China-Accessible Mirrors) ★ 70
How to Run an R&D Systems Quantikine ELISA
"Complete Quantikine workflow with hands-on benchwork and troubleshooting guidance is useful for correct technique and diagnosing reproducibility issues at curve extremes."
Source: abcam.com ↗
← Back to all cases