Home Biochemistry How to Run an R&D Systems Quantikine ELISA
Biochemistry Bilibili (China-Accessible Mirrors)

How to Run an R&D Systems Quantikine ELISA

🚨 Failure Case Library (41) + Submit your own case

critical
No Signal or Weak Signal in ELISA
No detectable signal or very weak signal across all wells, including standard wells. The plate reader shows absorbance values near baseline or significantly lower than expected.
💡 6 · ✓ 6
critical
No Signal or Weak Signal in ELISA
The ELISA plate reader detects no colorimetric signal or signal intensity is significantly below expected values after substrate development. Wells appear clear or only faintly colored even after standard incubation times of 10-30 minutes.
💡 6 · ✓ 6
critical
Poor or Non-Linear Standard Curve
Standard curve shows irregular shape, poor linearity, or inconsistent serial dilution response. R-squared value below 0.95 or curve does not follow expected sigmoidal or linear pattern across dynamic range.
💡 5 · ✓ 5
critical
Complete Absence of Standard Curve Signal with Normal Zero
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
💡 4 · ✓ 4
severe
B/B0 Ratio Outside Acceptable Range at Curve Endpoints
The B/B0 ratio (bound/maximum bound) for the terminal point of the standard curve is either >95% or <5%, indicating insufficient competitive displacement range.
💡 3 · ✓ 3
severe
Insufficient Signal from Zero Standard (B0)
The zero standard (no competing antigen, maximum antibody binding) produces OD values that are too low, resulting in compressed standard curve with poor sensitivity.
💡 4 · ✓ 4
severe
Standard Curve Plateau at High Concentration End
The standard curve reaches a plateau at high analyte concentrations (low OD values in competitive format), showing no further decrease in signal despite increasing standard concentration.
💡 4 · ✓ 4
severe
No or Very Low Signal Due to Poor Plate Binding
No detectable signal or extremely weak signal in ELISA even when target protein is expected to be present. Signal may be inconsistent across wells.
💡 4 · ✓ 4
severe
Primary and Secondary Antibody Incompatibility
No signal or minimal signal in indirect ELISA despite proper coating and blocking. Positive controls with matched antibodies work correctly.
💡 4 · ✓ 4
severe
Buffer Components Inhibit Detection Enzyme
No or very weak signal despite correct antibody binding. Signal loss occurs after addition of detection reagent. Other wells using different buffers show normal signal.
💡 4 · ✓ 4
severe
Antibody or Reagent Loss of Activity from Improper Storage
Progressive signal loss over time using the same reagents. Previously working protocols suddenly fail. Positive controls that worked before now show reduced or no signal.
💡 5 · ✓ 5
severe
Non-specific Secondary Antibody Binding Causing High Background
Elevated background signal across wells, including negative controls. Signal appears uniformly high rather than specific to target-containing wells.
💡 4 · ✓ 5
severe
Standard Curve R² Below 0.98 Threshold
The regression coefficient (R²) of the standard curve falls below 0.98, indicating poor curve fit and unreliable quantification of sample concentrations.
💡 5 · ✓ 6
severe
Poor Dynamic Range Between Signal and Background
The difference between maximum signal (high standards) and background (blank wells) is compressed, typically less than 5-fold, making it difficult to distinguish between samples of different concentrations.
💡 5 · ✓ 5
severe
Inadequate Washing Between Assay Steps
High background with retained signal in negative controls. Edge wells may show higher background than center wells. Residual reagents visible in wells.
💡 4 · ✓ 5
severe
Incomplete Standard Reconstitution with Visible Particulates
After adding reconstitution buffer to lyophilized standard, visible undissolved material or particulates remain in the vial, leading to inaccurate standard concentrations.
💡 4 · ✓ 6
severe
Inadequate Blocking of Non-Specific Binding Sites
Uniformly elevated background across plate with high coefficient of variation. Background may be particularly high in wells with lower antigen concentrations.
💡 4 · ✓ 4
severe
High Variability Between Replicates (CV >15%)
Replicate wells for the same sample or standard show high coefficient of variation (>15%), with inconsistent absorbance values that suggest uneven treatment or technical errors.
💡 5 · ✓ 5
severe
Uneven Color Development Across Plate
Color intensity varies significantly between wells that should be identical. Pattern may show edge effects, gradients across the plate, or random spotty appearance indicating incomplete or inconsistent reagent contact.
💡 4 · ✓ 4
severe
High Uniform Background Signal
All wells show elevated absorbance including blanks and negative controls, creating a uniformly high background that reduces signal-to-noise ratio and compresses the standard curve.
💡 5 · ✓ 5
severe
Poor Reproducibility Between Duplicate Wells
Coefficient of variation (CV) between duplicate or triplicate wells exceeds 10-15%. Individual replicates show inconsistent absorbance values that cannot be attributed to biological variation.
💡 4 · ✓ 4
moderate
Standard Curve Plateau at High Concentration End
The top of the standard curve (high analyte concentration, low OD) shows a plateau with minimal signal change across multiple high-concentration standards, reducing dynamic range.
💡 4 · ✓ 4
moderate
High Coefficient of Variation at Low Standard Concentrations
Replicate wells at the bottom of the standard curve (low analyte, high OD) show high CVs >15-20%, while mid-curve and high-concentration points demonstrate acceptable reproducibility.
💡 4 · ✓ 4
moderate
Wrong Instrument Settings for Detection Wavelength
No signal or very low signal readings from plate reader despite visible color development or expected fluorescence. Signal does not correlate with visual observations.
💡 4 · ✓ 4
moderate
Mixed Components from Different ELISA Kits
Unpredictable or absent signal despite following protocol. Results are inconsistent and do not match expected standard curves. Signal may vary dramatically between experiments.
💡 4 · ✓ 4
moderate
Incubation Temperature Below Optimal Range
Consistently weak signal across all samples including positive controls. Signal improves when experiment is repeated with attention to temperature. Longer incubations partially compensate.
💡 4 · ✓ 4
moderate
Over-Washing Removes Bound Detection Reagents
Signal progressively decreases with increased wash steps or aggressive washing. Reducing wash stringency restores signal. Edge wells show lower signal than center wells.
💡 4 · ✓ 5
moderate
Excessive Primary Antibody Concentration Saturating Wells
High background signal with poor signal-to-noise ratio. Standard curve may show compressed dynamic range or plateau at lower concentrations than expected.
💡 3 · ✓ 3
moderate
Excessive Substrate Concentration or Incubation Time
Rapid color development with all wells turning dark quickly. Signal may exceed linear range of reader (OD >2.0). Background wells show significant color development.
💡 4 · ✓ 5
moderate
Precipitate Formation in Wells Upon Substrate Addition
Visible precipitate or cloudiness appears in wells immediately or shortly after substrate addition. Absorbance readings may be abnormally high or variable.
💡 4 · ✓ 4
moderate
Excessive Signal Amplification in Detection System
Very high signals across all wells including low-concentration standards. Loss of discrimination between different antigen concentrations. Background approaches signal levels.
💡 4 · ✓ 4
moderate
Biological Sample Not in Detectable Range
Standard curve appears normal, but sample wells show no signal or signal below the lowest standard. This occurs when the analyte concentration in samples is below the assay detection limit.
💡 4 · ✓ 4
moderate
High Variability Between Experimental Runs
Standard curves or sample values differ significantly between experiments run on different days, despite using the same reagents and protocol, showing poor reproducibility.
💡 5 · ✓ 5
moderate
Edge and Drift Effects on ELISA Plate
Wells at the plate edges show systematically higher or lower absorbance than center wells, or a gradient pattern appears across the plate, affecting data quality especially for samples in edge positions.
💡 5 · ✓ 5
moderate
Weak Signal from Inadequate Antibody-Antigen Interaction
Signal is present but consistently lower than expected. Positive controls show weak but detectable signal while samples are near background.
💡 4 · ✓ 4
moderate
Standard Curve Fitting Model Not Appropriate
The recommended curve-fitting model (typically Four-Parameter Logistic) does not adequately fit the standard data points, resulting in poor correlation even with properly prepared standards.
💡 4 · ✓ 5
moderate
Bacterial Contamination in Wash or Incubation Buffers
Sporadic signal loss or high background that worsens over time during multi-day experiments. Cloudy or turbid buffer appearance. Inconsistent results when using same buffer batch.
💡 4 · ✓ 5
moderate
High Coefficient of Variation Between Standard Replicates
Duplicate or triplicate wells for the same standard concentration show high variability (CV >10-15%), compromising curve fit and confidence in quantification.
💡 5 · ✓ 6
minor
Standard Curve OD Values Differ From Datasheet
The optical density (OD) measurement values for the standard curve vary considerably from the examples shown on the kit datasheet or protocol booklet, causing user concern about assay validity.
💡 4 · ✓ 4
minor
Sample Values Above Assay Range (Hook Effect Excluded)
Sample absorbance readings exceed the top standard, reading off the curve, while the standard curve itself appears normal with proper shape and dynamic range.
💡 2 · ✓ 3
minor
Green Color Upon Adding Stop Solution (Streptavidin-HRP)
When sulfuric acid stop solution is added to wells after TMB substrate incubation, a green color appears instead of the expected yellow, indicating incomplete mixing and pH gradient in the well.
💡 2 · ✓ 3
💬 Comments coming soon