Home Failure Case Library Complete Absence of Standard Curve Signal with Normal Zero
ELISA (Competitive) critical

Complete Absence of Standard Curve Signal with Normal Zero

Symptom
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
Common Causes
  1. 1 Pipetting error causing omission of standards from wells during plate setup
  2. 2 Serial dilution error resulting in incorrect standard concentrations
  3. 3 Standard concentration vastly exceeds assay dynamic range causing 100% competition at all points
  4. 4 Standard stock solution degraded or incorrectly prepared
Solutions
  1. 1 Verify standard curve pipetting protocol with careful visual confirmation of additions to each well
  2. 2 Recalculate and re-prepare serial dilutions using fresh standard stock solution
  3. 3 Decrease starting standard concentration by 10-100 fold to bring into assay dynamic range
  4. 4 Confirm standard stock concentration and activity using alternative analytical method (mass spec, HPLC)
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
ELISA Operation Demonstration Video
"Concise hands-on demonstration of proper ELISA pipetting and plate setup, directly addressing the top cause of omitted standards."
Bilibili (China-Accessible Mirrors) ★ 80
ELISA Basic Operation Steps Demonstration
"Step-by-step basic operation video emphasizes correct well dispensing, useful for avoiding pipetting errors that cause missing standard signal."
Thermo Fisher Scientific ★ 75
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Tutorial on running an ELISA kit covers accurate reagent addition, relevant to preventing standard omission during plate preparation."
Source: abcam.com ↗
← Back to all cases