Home Immunology ELISA Experimental Protocol Complete Series (5 Episodes)
Immunology Bilibili (China-Accessible Mirrors)

ELISA Experimental Protocol Complete Series (5 Episodes)

🚨 Failure Case Library (9) + Submit your own case

critical
Poor or Non-Linear Standard Curve
Standard curve shows irregular shape, poor linearity, or inconsistent serial dilution response. R-squared value below 0.95 or curve does not follow expected sigmoidal or linear pattern across dynamic range.
💡 5 · ✓ 5
severe
Detection System Lacks Sensitivity for Target Concentration
Low or no signal despite proper assay setup. High concentration samples may show weak signal while low concentration samples are undetectable.
💡 4 · ✓ 4
severe
Non-specific Secondary Antibody Binding Causing High Background
Elevated background signal across wells, including negative controls. Signal appears uniformly high rather than specific to target-containing wells.
💡 4 · ✓ 5
severe
Serial Dilution Calculation or Execution Error
Standard curve shows irregular pattern or non-monotonic response, with individual points deviating significantly from expected sigmoidal or linear trend.
💡 5 · ✓ 6
severe
Inadequate Washing Between Assay Steps
High background with retained signal in negative controls. Edge wells may show higher background than center wells. Residual reagents visible in wells.
💡 4 · ✓ 5
severe
High CV from Inconsistent Well Washing
High coefficient of variation (>15%) between replicates with uneven background signal across the plate. Some wells show higher background than others after washing steps.
💡 5 · ✓ 5
moderate
Standard Curve Plateau at High Concentration End
The top of the standard curve (high analyte concentration, low OD) shows a plateau with minimal signal change across multiple high-concentration standards, reducing dynamic range.
💡 4 · ✓ 4
moderate
Excessive Primary Antibody Concentration Saturating Wells
High background signal with poor signal-to-noise ratio. Standard curve may show compressed dynamic range or plateau at lower concentrations than expected.
💡 3 · ✓ 3
moderate
B/B0 Ratio at Standard Curve Endpoint Out of Range
The B/B0 ratio (bound/maximum bound) at the lowest or highest standard concentration is either >95% or <5%, indicating the standard curve dynamic range is not properly positioned.
💡 3 · ✓ 3