Home Failure Case Library Inadequate Blocking of Non-Specific Binding Sites
ELISA (High Background) severe

Inadequate Blocking of Non-Specific Binding Sites

Symptom
Uniformly elevated background across plate with high coefficient of variation. Background may be particularly high in wells with lower antigen concentrations.
Common Causes
  1. 1 Blocking incubation time insufficient to saturate all binding sites
  2. 2 Blocking agent incompatible with assay system or antibodies
  3. 3 Blocking protein concentration too low
  4. 4 Blocking buffer does not match detection system species
Solutions
  1. 1 Increase blocking incubation period (typically 1-2 hours at room temperature)
  2. 2 Use 5-10% normal serum from same species as detection antibody
  3. 3 Test alternative blocking agents (BSA, casein, non-fat dry milk)
  4. 4 Ensure blocking buffer concentration is adequate (typically 3-5% protein)
Related Video (3)