R&D Systems Quantikine ELISA Operation Guide

🚨 Failure Case Library (41) + Submit your own case

critical
Absent Standard Curve Signal with Normal Zero Standard
No detectable signal is observed for standard curve points containing analyte, while the zero standard (maximum binding control) shows normal OD values, indicating complete loss of competitive inhibition.
💡 4 · ✓ 4
critical
No Signal or Weak Signal in ELISA
After developing the ELISA plate, wells show no color change or very faint signal that is barely above background, even for positive controls or high-concentration standards.
💡 6 · ✓ 6
critical
No Signal or Weak Signal in ELISA
No detectable signal or very weak signal across all wells, including standard wells. The plate reader shows absorbance values near baseline or significantly lower than expected.
💡 6 · ✓ 6
critical
Poor or Non-Linear Standard Curve
Standard curve shows irregular shape, poor linearity, or inconsistent serial dilution response. R-squared value below 0.95 or curve does not follow expected sigmoidal or linear pattern across dynamic range.
💡 5 · ✓ 5
critical
No Signal for Standard Curve with Normal Zero Standard
All standard concentrations show no signal (high OD in competitive format similar to zero standard), while the zero standard (no competitor) shows normal maximum binding signal.
💡 4 · ✓ 4
critical
Complete Absence of Standard Curve Signal with Normal Zero
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
💡 4 · ✓ 4
severe
Excessive Non-Specific Background Signal in Competitive ELISA
High background optical density readings are observed across wells, obscuring the difference between standards and samples. Signal persists even in wells without primary antibody or sample.
💡 4 · ✓ 4
severe
B/B0 Ratio Outside Acceptable Range at Curve Endpoints
The B/B0 ratio (bound/maximum bound) for the terminal point of the standard curve is either >95% or <5%, indicating insufficient competitive displacement range.
💡 3 · ✓ 3
severe
Standard Curve Plateau at High Concentration Range
The standard curve reaches a plateau (flat response) at high standard concentrations with correspondingly low OD values, failing to show expected dose-response relationship in competitive ELISA format.
💡 4 · ✓ 4
severe
Non-specific Background Signal Too High in Competitive ELISA
High background optical density is observed across wells, reducing signal-to-noise ratio and making it difficult to distinguish specific binding from non-specific interactions.
💡 4 · ✓ 4
severe
Standard Curve Plateau at High Concentration End
The standard curve reaches a plateau at high analyte concentrations (low OD values in competitive format), showing no further decrease in signal despite increasing standard concentration.
💡 4 · ✓ 4
severe
OD Value of Zero Standard Too Low
The zero standard (maximum binding, no competitor) shows insufficient optical density signal, providing inadequate dynamic range for the competitive assay and poor B/B0 calculations.
💡 4 · ✓ 4
severe
Antibody or Reagent Loss of Activity from Improper Storage
Progressive signal loss over time using the same reagents. Previously working protocols suddenly fail. Positive controls that worked before now show reduced or no signal.
💡 5 · ✓ 5
severe
Non-specific Secondary Antibody Binding Causing High Background
Elevated background signal across wells, including negative controls. Signal appears uniformly high rather than specific to target-containing wells.
💡 4 · ✓ 5
severe
High CV from Inconsistent Pipetting Technique
High intra-assay coefficient of variation (>15%) between technical replicates. Standard curve shows poor reproducibility between duplicate points.
💡 5 · ✓ 5
severe
High CV from Plate Edge Effects
Outer (edge) wells show systematically different absorbance values compared to inner wells, with CV >15%. Edge wells may show higher or lower signal depending on evaporation or temperature variation. Visual inspection from the side reveals lower buffer levels in edge wells.
💡 5 · ✓ 5
severe
Standard Curve R² Below 0.98 Threshold
The regression coefficient (R²) of the standard curve falls below 0.98, indicating poor curve fit and unreliable quantification of sample concentrations.
💡 5 · ✓ 6
severe
Serial Dilution Calculation or Execution Error
Standard curve shows irregular pattern or non-monotonic response, with individual points deviating significantly from expected sigmoidal or linear trend.
💡 5 · ✓ 6
severe
B/B0 Ratio at Curve Endpoint Outside 5-95% Range
The B/B0 ratio (bound/maximum bound signal) for the highest or lowest standard concentration exceeds 95% or falls below 5%, indicating the standard curve dynamic range does not adequately cover the analyte concentration range.
💡 4 · ✓ 4
severe
High Non-Specific Background Signal in Competitive ELISA
Elevated background signal is detected across wells, interfering with specific signal measurement. Background noise reduces assay sensitivity and may mask low-concentration analyte detection.
💡 4 · ✓ 4
severe
Inadequate Washing Between Assay Steps
High background with retained signal in negative controls. Edge wells may show higher background than center wells. Residual reagents visible in wells.
💡 4 · ✓ 5
severe
High Inter-Assay CV from Inconsistent Sample Storage
High inter-assay coefficient of variation (>15%) when the same samples are tested across different runs. Sample readings decrease over time or vary unpredictably between runs.
💡 5 · ✓ 5
severe
Incomplete Standard Reconstitution with Visible Particulates
After adding reconstitution buffer to lyophilized standard, visible undissolved material or particulates remain in the vial, leading to inaccurate standard concentrations.
💡 4 · ✓ 6
severe
Low OD Value for Zero Standard in Competitive ELISA
The zero standard (B0, representing maximum antibody binding without competing analyte) produces lower than expected optical density values, reducing overall assay sensitivity and dynamic range.
💡 4 · ✓ 4
severe
High CV from Inconsistent Well Washing
High coefficient of variation (>15%) between replicates with uneven background signal across the plate. Some wells show higher background than others after washing steps.
💡 5 · ✓ 5
severe
Inadequate Blocking of Non-Specific Binding Sites
Uniformly elevated background across plate with high coefficient of variation. Background may be particularly high in wells with lower antigen concentrations.
💡 4 · ✓ 4
severe
High Uniform Background Signal
All wells show elevated absorbance including blanks and negative controls, creating a uniformly high background that reduces signal-to-noise ratio and compresses the standard curve.
💡 5 · ✓ 5
severe
High Background Signal in ELISA
All wells including negative controls show elevated absorbance readings. Background signal obscures specific signal, reducing signal-to-noise ratio and making data interpretation difficult.
💡 6 · ✓ 6
moderate
Standard Curve Plateau at High Concentration End
The top of the standard curve (high analyte concentration, low OD) shows a plateau with minimal signal change across multiple high-concentration standards, reducing dynamic range.
💡 4 · ✓ 4
moderate
Mixed Components from Different ELISA Kits
Unpredictable or absent signal despite following protocol. Results are inconsistent and do not match expected standard curves. Signal may vary dramatically between experiments.
💡 4 · ✓ 4
moderate
Over-Washing Removes Bound Detection Reagents
Signal progressively decreases with increased wash steps or aggressive washing. Reducing wash stringency restores signal. Edge wells show lower signal than center wells.
💡 4 · ✓ 5
moderate
Precipitate Formation in Wells Upon Substrate Addition
Visible precipitate or cloudiness appears in wells immediately or shortly after substrate addition. Absorbance readings may be abnormally high or variable.
💡 4 · ✓ 4
moderate
Unexpected or Inconsistent Assay Results
Sample quantification values are outside expected biological range, controls fail to meet acceptance criteria, or results are not reproducible between runs despite similar experimental conditions.
💡 5 · ✓ 5
moderate
Biological Sample Not in Detectable Range
Standard curve appears normal, but sample wells show no signal or signal below the lowest standard. This occurs when the analyte concentration in samples is below the assay detection limit.
💡 4 · ✓ 4
moderate
High Variability Between Experimental Runs
Standard curves or sample values differ significantly between experiments run on different days, despite using the same reagents and protocol, showing poor reproducibility.
💡 5 · ✓ 5
moderate
High Coefficient of Variation at Lower Standard Curve
Replicate measurements at the bottom of the standard curve (highest OD values, lowest analyte concentrations) show excessive variability with high coefficient of variation (CV) between replicates.
💡 4 · ✓ 4
moderate
Bacterial Contamination in Wash or Incubation Buffers
Sporadic signal loss or high background that worsens over time during multi-day experiments. Cloudy or turbid buffer appearance. Inconsistent results when using same buffer batch.
💡 4 · ✓ 5
moderate
High CVs at Bottom of Standard Curve
Coefficient of variation (CV) among replicates is excessively high at the lowest standard concentrations (highest OD in competitive format), reducing precision at the sensitive end of the curve.
💡 4 · ✓ 4
moderate
High Coefficient of Variation Between Standard Replicates
Duplicate or triplicate wells for the same standard concentration show high variability (CV >10-15%), compromising curve fit and confidence in quantification.
💡 5 · ✓ 6
minor
Sample Values Above Assay Range (Hook Effect Excluded)
Sample absorbance readings exceed the top standard, reading off the curve, while the standard curve itself appears normal with proper shape and dynamic range.
💡 2 · ✓ 3
minor
Green Color Upon Adding Stop Solution (Streptavidin-HRP)
When sulfuric acid stop solution is added to wells after TMB substrate incubation, a green color appears instead of the expected yellow, indicating incomplete mixing and pH gradient in the well.
💡 2 · ✓ 3