Home › Failure Case Library › High Inter-Assay CV from Inconsistent Sample Storage
ELISA (Inconsistent Results / High CV) severe

High Inter-Assay CV from Inconsistent Sample Storage

Symptom
High inter-assay coefficient of variation (>15%) when the same samples are tested across different runs. Sample readings decrease over time or vary unpredictably between runs.
Common Causes
  1. 1 Samples stored at inconsistent temperatures between runs
  2. 2 Unequal number of freeze-thaw cycles across different samples
  3. 3 Sample preparation methods vary between runs (dilution, mixing, timing)
  4. 4 Samples not stored according to manufacturer recommendations (temperature, additives)
  5. 5 Degradation of analyte in samples due to improper storage conditions
Solutions
  1. 1 Ensure all samples are stored and prepared consistently according to kit recommendations
  2. 2 Keep the number of freeze-thaw cycles equal across all samples; aliquot samples to avoid repeated thawing
  3. 3 Document and follow standardized sample preparation protocols for all runs
  4. 4 Store samples at recommended temperature (typically -20°C or -80°C for long-term storage)
  5. 5 Add protease inhibitors or stabilizers if recommended by the protocol
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 78
R&D Systems Quantikine ELISA Operation Guide
"R&D Systems Quantikine guide covers sample preparation through detection, the workflow where storage/temperature inconsistency causes inter-assay CV."
Thermo Fisher Scientific ★ 72
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Step-by-step Invitrogen tutorial emphasizes accurate, reproducible protein detection, directly contextualizing reproducibility failure modes like high inter-assay CV."
Bilibili (China-Accessible Mirrors) ★ 68
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"DuoSet hands-on protocol includes sample preparation and full sandwich ELISA steps where storage handling affects assay variability."
Source: abcam.com ↗
← Back to all cases