No Signal or Weak Signal in ELISA

Symptom
No detectable signal or very weak signal across all wells, including standard wells. The plate reader shows absorbance values near baseline or significantly lower than expected.
Common Causes
  1. 1 Reagents added in incorrect order or prepared incorrectly
  2. 2 Antibody concentration too low (primary or secondary)
  3. 3 Primary and secondary antibodies not compatible (species mismatch)
  4. 4 Capture antibody or antigen did not adhere to plate (wrong plate type or insufficient coating time)
  5. 5 Capture and detection antibodies recognize the same epitope in sandwich ELISA
  6. 6 Sodium azide in buffers inhibiting HRP activity
Solutions
  1. 1 Repeat experiment following protocol exactly for solution preparation and addition order
  2. 2 Increase antibody concentration through titration; extend incubation to overnight at 4°C
  3. 3 Verify secondary antibody was raised against primary antibody host species (e.g. anti-mouse for mouse primary)
  4. 4 Use ELISA-validated plates (not tissue culture plates); extend coating step to overnight at 4°C
  5. 5 Verify antibodies recognize distinct epitopes; use validated matched antibody pairs
  6. 6 Use sodium azide-free buffers or ensure sufficient washing to remove azide completely
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
How to Run an R&D Systems Quantikine ELISA
"Demonstrates complete R&D Systems Quantikine ELISA workflow with hands-on benchwork and troubleshooting guidance, directly addressing reagent addition sequence and proper protocol execution to prevent"
Bilibili (China-Accessible Mirrors) ★ 82
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"Shows step-by-step sandwich ELISA procedure including critical plate coating, antibody addition sequence, and substrate development stages where reagent order errors commonly cause signal loss."
Bilibili (China-Accessible Mirrors) ★ 78
R&D Systems Quantikine ELISA Operation Guide
"Official R&D Systems Quantikine ELISA protocol demonstration covering complete benchwork procedures from sample preparation through detection, ensuring proper reagent preparation and addition order."
Source: rndsystems.com ↗
← Back to all cases