Sample Values Above Assay Range (Hook Effect Excluded)

Symptom
Sample absorbance readings exceed the top standard, reading off the curve, while the standard curve itself appears normal with proper shape and dynamic range.
Common Causes
  1. 1 Sample contains analyte (cytokine, protein) at concentration above the upper limit of quantification (ULOQ)
  2. 2 Sample not diluted or dilution factor insufficient for analyte concentration
Solutions
  1. 1 Dilute samples to bring values within standard curve range (typically 1:5 to 1:100 depending on analyte)
  2. 2 Re-run samples with serial dilutions (e.g. 1:2, 1:10, 1:50) to identify optimal dilution
  3. 3 Multiply final result by dilution factor to calculate actual concentration
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
How to Run an R&D Systems Quantikine ELISA
"Demonstrates complete R&D Systems Quantikine ELISA workflow with troubleshooting guidance, directly relevant to understanding how proper standard curves are generated and where sample overrange issues"
Bilibili (China-Accessible Mirrors) ★ 82
R&D Systems Quantikine ELISA Operation Guide
"Official R&D Systems Quantikine ELISA protocol showing sample preparation through detection, essential for understanding correct assay execution and recognizing when samples exceed the quantifiable ra"
Bilibili (China-Accessible Mirrors) ★ 72
Thermo Fisher Microplate Reader Software Operation
"Covers microplate reader software operation and data acquisition, directly relevant to understanding how absorbance readings are generated and interpreted when values exceed the standard curve."
Source: rndsystems.com ↗
← Back to all cases