Low RNA Yield After Cleanup

Symptom
Recovered RNA concentration is significantly lower than expected based on input amount. Spectrophotometric measurement shows inadequate RNA quantity for downstream applications.
Common Causes
  1. 1 Incorrect buffer reconstitution or order of reagent addition during protocol execution
  2. 2 Insufficient mixing of ethanol with RNA sample and RNA Cleanup Binding Buffer before column loading
  3. 3 Incomplete saturation of column matrix - elution water not delivered directly to center of column
  4. 4 High degree of RNA secondary structure affecting binding and elution of small RNAs (< 45 nt)
  5. 5 Premature discarding of flow-through or eluents during protocol steps
Solutions
  1. 1 Verify protocol adherence: confirm correct buffer reconstitution, reagent addition order, and proper handling of all flow-through and eluent fractions
  2. 2 Thoroughly mix ethanol with RNA sample and RNA Cleanup Binding Buffer by vortexing or pipetting before applying to column
  3. 3 Deliver nuclease-free elution water directly to the center of the column matrix; allow complete saturation before centrifugation
  4. 4 For small RNAs (< 45 nt) with suspected secondary structure, dilute sample with 2 volumes of ethanol instead of 1 volume in Step 2
  5. 5 Use larger elution volumes (note: will dilute sample), perform multiple elutions, or extend incubation time to 2-5 minutes on column before centrifugation
Related Video (2)
Bilibili (China-Accessible Mirrors) ★ 75
Trizol RNA Extraction from Cells: Hands-On Protocol
"Covers Trizol RNA extraction with troubleshooting tips that may address buffer and reagent issues."
Bilibili (China-Accessible Mirrors) ★ 70
Trizol RNA Extraction: Protocol & Principles
"Combines hands-on protocol with mechanistic explanation, helping understand reagent order and separation chemistry."
Source: neb.com ↗
← Back to all cases