Poor or Non-Linear Standard Curve

Symptom
Standard curve shows irregular shape, poor linearity, or inconsistent serial dilution response. R-squared value below 0.95 or curve does not follow expected sigmoidal or linear pattern across dynamic range.
Common Causes
  1. 1 Pipetting errors or calculation mistakes in preparing serial dilution series
  2. 2 Poor or variable adsorption of coating reagents to plate surface
  3. 3 Capture antibody did not properly bind to plate due to incorrect coating buffer or presence of interfering proteins
  4. 4 Reagents poorly mixed before dispensing into wells
  5. 5 Inhomogeneous samples or improper ELISA plate selection
Solutions
  1. 1 Verify pipetting technique and dilution calculations; use calibrated pipettes; prepare fresh dilution series
  2. 2 Use proper coating buffer: PBS pH 7.4 or carbonate-bicarbonate buffer pH 9.6; extend coating incubation time if needed
  3. 3 Dilute capture antibody in PBS without other proteins; verify proper high-binding ELISA plate is used
  4. 4 Thoroughly mix reagents before use; vortex or invert multiple times
  5. 5 Check sample homogeneity; consider different plate types (high-binding vs medium-binding); use appropriate plates for fluorescence detection if applicable
Related Video (2)
Bilibili (China-Accessible Mirrors) ★ 78
ELISA Experimental Protocol Complete Series (5 Episodes)
"Comprehensive multi-episode ELISA benchwork series likely covers serial dilution and standard curve setup, directly addressing the top cause."
Bilibili (China-Accessible Mirrors) ★ 72
How to Run an R&D Systems Quantikine ELISA
"Quantikine workflow video includes hands-on techniques and troubleshooting guidance, useful for avoiding dilution/pipetting errors affecting standard curves."
Source: sigmaaldrich.com ↗
← Back to all cases