Amplification Plots Dip Below Zero Due to Incorrect Baseline Settings

Symptom
Amplification plots are clearly abnormal with sections dipping below zero dR; data cannot be used as presented; plots appear distorted
Common Causes
  1. 1 Inappropriate baseline settings applied by instrument software
  2. 2 Automatic baseline detection failed
  3. 3 Baseline cycles incorrectly defined
Solutions
  1. 1 Examine raw data plot to verify actual amplification profile is normal
  2. 2 Manually set appropriate baseline cycles based on raw data (typically the initial linear, baseline phase of the reaction, e.g., cycles 6 to 16)
  3. 3 Reanalyze data with corrected baseline settings
  4. 4 Consult instrument manual for proper baseline determination