Home Failure Case Library Amplification Plots Dip Below Zero Due to Incorrect Baseline Settings
PCR (Sigma Guide) moderate

Amplification Plots Dip Below Zero Due to Incorrect Baseline Settings

Symptom
Amplification plots are clearly abnormal with sections dipping below zero dR; data cannot be used as presented; plots appear distorted
Common Causes
  1. 1 Inappropriate baseline settings applied by instrument software
  2. 2 Automatic baseline detection failed
  3. 3 Baseline cycles incorrectly defined
Solutions
  1. 1 Examine raw data plot to verify actual amplification profile is normal
  2. 2 Manually set appropriate baseline cycles based on raw data (typically the initial linear, baseline phase of the reaction, e.g., cycles 6 to 16)
  3. 3 Reanalyze data with corrected baseline settings
  4. 4 Consult instrument manual for proper baseline determination