Home › Failure Case Library › No Signal for Standard Curve with Normal Zero Standard
ELISA (Competitive) critical

No Signal for Standard Curve with Normal Zero Standard

Symptom
All standard concentrations show no signal (high OD in competitive format similar to zero standard), while the zero standard (no competitor) shows normal maximum binding signal.
Common Causes
  1. 1 Standard curve pipetting errors resulting in no analyte addition
  2. 2 Incorrect serial dilution procedure producing uniform zero concentration
  3. 3 Standard concentration far exceeds assay dynamic range causing complete inhibition at all points
  4. 4 Standard stock solution degraded or improperly reconstituted
Solutions
  1. 1 Verify the standard curve was pipetted and diluted properly with documented procedure
  2. 2 Check calculation of dilution factors and physically verify volumes
  3. 3 Decrease the starting standard concentration to fall within assay dynamic range
  4. 4 Prepare fresh standard from new stock and verify reconstitution protocol
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 72
R&D Systems Quantikine ELISA Operation Guide
"Official R&D Systems Quantikine guide demonstrates the full competitive/sandwich ELISA workflow including standard and sample preparation, directly showing where pipetting errors cause the no-signal c"
Bilibili (China-Accessible Mirrors) ★ 70
How to Run an R&D Systems Quantikine ELISA
"Quantikine protocol video explicitly includes troubleshooting guidance, useful for understanding standard-curve pipetting and signal issues."
Thermo Fisher Scientific ★ 65
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Invictrogen step-by-step kit tutorial shows correct standard/sample loading and plate setup, helping a researcher avoid the pipetting error behind the failed curve."
Source: abcam.com ↗
← Back to all cases