Home Failure Case Library Unexpected or Inconsistent Assay Results
ELISA (Sigma Guide) moderate

Unexpected or Inconsistent Assay Results

Symptom
Sample quantification values are outside expected biological range, controls fail to meet acceptance criteria, or results are not reproducible between runs despite similar experimental conditions.
Common Causes
  1. 1 Omission of critical reagents or incorrect preparation of working solutions
  2. 2 Dilution errors in sample or standard preparation
  3. 3 Technique problems with reagent mixing or washing procedures
  4. 4 Inappropriate ELISA plate used (e.g. non-fluorescence compatible plates for fluorescence detection)
  5. 5 Blocking protein inadvertently included in coating solution, preventing proper antigen/antibody binding
Solutions
  1. 1 Verify reagents prepared correctly following manufacturer instructions; use preparation checklist; confirm reagents added in correct order
  2. 2 Double-check pipetting technique and all dilution calculations; use calibrated pipettes
  3. 3 Ensure proper mixing of all reagents before use; standardize wash step protocols with consistent timing and volumes
  4. 4 Use appropriate plate type for detection method: fluorescence-compatible black or white plates for fluorescence; clear plates for colorimetric; verify plate specifications
  5. 5 Omit blocking proteins (BSA, casein, milk) from coating solution; add blocking step only after coating and washing
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
R&D Systems Quantikine ELISA Operation Guide
"Official R&D Systems Quantikine protocol covers complete benchwork including sample preparation and reagent handling, directly addressing reagent omission and working solution preparation errors."
Bilibili (China-Accessible Mirrors) ★ 82
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"Bio-Techne DuoSet hands-on demonstration explicitly shows plate coating, capture/detection antibody steps, and substrate development—critical reagent preparation stages where omissions cause control f"
Thermo Fisher Scientific ★ 78
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Invitrogen kit tutorial focuses on achieving accurate and reproducible results with explicit protocol steps, directly relevant to diagnosing reproducibility failures from incorrect working solution pr"
Source: sigmaaldrich.com ↗
← Back to all cases